• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI SPECTRO2 30 PULSE OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI SPECTRO2 30 PULSE OXIMETER Back to Search Results
Model Number WW1030EN
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the device rejected by pulmolink at goods inwards acceptance testing displaying "10 err".No patient injury reported.
 
Additional Manufacturer Narrative
This device returned out of box and determined the failure resulted from intermittent connection with the main board processor caused by suspected improper solder joint or crack in the trace.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.Device manufactured in 2020-12.Operator of device is unknown.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SPECTRO2 30 PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15326463
Report Number3012307300-2022-16905
Device Sequence Number12097196
Product Code DQA
UDI-Device Identifier30843418000191
UDI-Public30843418000191
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K083557
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberWW1030EN
Device Catalogue NumberWW1030EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/15/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/10/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-