ST PAUL BCI SMITHS MEDICAL SPECTRO2 PULSE OXIMETER Back to Search Results
Model Number WW1000A1EN
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the suspect device was providing incorrect reading.No patient injury was reported.
 
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated that all inspections were completed and the manufacturing process at the manufacturing facility and the suspect device manufacturing records are retained by the supplier at their manufacturing site and are not readily available for review.The suspect device was returned for evaluation.Physical and functional testing was performed and found that the issue was an out of box failure.Several tests were performed including; connecting the suspect device to a known good testing probe and the reported issue was duplicated and confirmed.Further evaluation determined that when flapping the clips constantly in an open and closed movement, the advisor monitor consistently detects and displays a heart rate.Evidence of noise interfering with the emitter and the detector signal which contributed to the customer's observed failure of erratic and erroneous readings.Concluded that the detector in the suspected device may have been malfunctioning all this time.However, there was no difference in the form and fit of the returned suspect device as alleged by the customer.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SMITHS MEDICAL SPECTRO2 PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15326568
Report Number3012307300-2022-16915
Device Sequence Number6484414
Product Code DQA
UDI-Device Identifier30843418000283
UDI-Public30843418000283
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K083787
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWW1000A1EN
Device Catalogue NumberWW1000A1EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1