• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GLOBUS MEDICAL, INC. CREO MIS; CREO MIS LOCKING CAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GLOBUS MEDICAL, INC. CREO MIS; CREO MIS LOCKING CAP Back to Search Results
Model Number 1134.0010
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Failure of Implant (1924)
Event Date 07/25/2022
Event Type  malfunction  
Manufacturer Narrative
Three locking caps were returned for evaluation.Additional information provided that the position and size of the right and left pedicles were different due to congenital malformation.It is possible that the congential malformation of the pedicles resulted in poor anchoring of the screw within the l4 pedicle.However, an exact cause of the reported issue could not be determined.
 
Event Description
It was reported that a revision surgery was needed to replace (3) creo locking caps that caused subsequent screw migration from l5 post operatively.This event occurred in japan.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CREO MIS
Type of Device
CREO MIS LOCKING CAP
Manufacturer (Section D)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer (Section G)
GLOBUS MEDICAL, INC.
2560 general armistead avenue
audubon PA 19403
Manufacturer Contact
daniel paul
2560 general armistead avenue
audubon, PA 19403
6109301800
MDR Report Key15326573
MDR Text Key305418190
Report Number3004142400-2022-00124
Device Sequence Number1
Product Code MNH
UDI-Device Identifier00889095071139
UDI-Public00889095071139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K124058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1134.0010
Device Lot NumberBAA030WC
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2022
Date Manufacturer Received08/03/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexFemale
Patient RaceAsian
-
-