| Model Number |
WW1030EN |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This unit returned to service less than a month after purchase and the reported problem could not be confirmed.Performed spo2 simulation test and monitor reads properly; spo2 also tested using the returned sensor finger probe and reads properly.Open up the monitor and did not find anything wrong with the monitor.A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|
| |
|
Event or Problem Description
|
|
It was reported that the device is presenting with sensor error.The sensors were changed and the device still gave an error.When the sensor were used on different device, they operated properly.No patient injury was reported.
|
| |
|