Model Number WW1000A1EN
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device and ear probes were received for evaluation.Visual and functional testing were performed: the device was in good condition with no physical damage or defects noted; the reported erratic reading was confirmed and isolated to the ear probes.The oximeter was found to be operating as intended and acquired readings accurately within specifications when tested on a 4086 multiparameter simulator and when finger sensors were used.The root cause of the reported issue was found to be a supplier issue.Malfunctioning ear probes as a result of detector sensitivity to noises and causing erratic readings.The failure resulted from malfunctioning accessories (ear probes) shipped with the monitor.A device history record (dhr) review is not applicable because the manufacturing records are retained by the supplier at their manufacturing site and are not readily available for review.Additional information has been requested from the supplier.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that, with use of the device, the size of the ear probe was too big, proved difficult to position properly, and the readings were very erratic.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI WW1000 PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15326719
Report Number3012307300-2022-16933
Device Sequence Number17477604
Product Code DQA
UDI-Device Identifier30843418000283
UDI-Public30843418000283
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K083787
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWW1000A1EN
Device Catalogue NumberWW1000A1EN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/17/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-