| Model Number |
WW1030EN |
| Medical Device Problem Code |
Device Alarm System (1012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device first time in for service after 1 year in the field and the reported problem is a user perception resulting from user interfacing with the unit in a manner inconsistent with the instructions for use (ifu).It is a normal operation of the spectro2 to alert the customer when no reading is being acquired or no patient is being detected because a patient is not attached.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.Lot # and initial reporter also sent report to fda are unknown.No information provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the device was displaying error message warranty.No patient injury reported.
|
| |
|
Search Alerts/Recalls
|