Model Number 3301A1
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/1901
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device first time in for service after 8 years in the field and the failure resulted from malfunctioning main board.The device history record (dhr) review is not applicable or relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.Operator of device and initial reporter also sent to fda?: are unknown.No information provided to date.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that the device is presenting with an error.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI SPECTRO2 PULSE OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15327077
Report Number3012307300-2022-16977
Device Sequence Number6361117
Product Code DQA
UDI-Device Identifier30843418000849
UDI-Public30843418000849
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K943404
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number3301A1
Device Catalogue Number3301A1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/14/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2012
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-