• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL DELTEC; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL DELTEC; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-2965-24
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.The returned samples were received decontaminated with its original packaging.Visual and functional testing were performed.The samples were tested with a syringe to replicate the failure mode.No leaks or air were detected in the returned samples.The complaint was not confirmed.Quality personnel takes several samples from loading before the material is packed.Perform a random sampling twice per shift or at least once per lot if lot is less than half a shift in order to verify that flow test is properly performed.Root cause cannot be determined as the complaint was not confirmed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Event or Problem Description
It was reported that air was noticed when pulling back into the syringe.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15327348
Report Number3012307300-2022-16993
Device Sequence Number6361628
Product Code FPA
UDI-Device Identifier10610586027734
UDI-Public10610586027734
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K021999
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/15/2025
Device Model Number21-2965-24
Device Catalogue Number21-2965-24
Device Lot Number4002680
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/18/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/22/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-