| Model Number |
21587600 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The actual device has passed both pcba test and final device test.Root cause analysis done in trended case states- by visual inspection it is observed that the surroundings of the usb interface has been deformed by heating.The usb connector in the device, as well as the connector on the cable are further investigated with a microscope.Functional testing reveals that the charger works fine, when running on the internal battery.That means that the malfunction in the device is purly a breakdown of the micro usb connector.The connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.Further information will be submitted as we get more follow up.
|
| |
|
Event or Problem Description
|
|
Sf (b)(4).Re charger- hot and melted.It was reported- hcp called tech support and reported that pt's charger got hot and melted at the junction of the charging cord and charger.Issue occurred on june 6th.Reportedly a new charger has already been sent to pt, but hcp was advised to call to report the issue.He said the charger with the issue has not been received yet, but he will send it to us.Hcp declined harm to pt.He read the notes from another provider that indicated the charger is still functioning, but damaged and occurred 1 time.Pt thinks a possible power surge.I advised to stop using the damaged charger.Hcp is unsure if has were in charger when the issue occurred.If has were in charger when the issue occurred, i advised to send the has in for evaluation as well.Qe form emailed to hcp.Flagged in navision.Complaint created.No harm was made to the user.This is a late case.This is an initial report.Further information will be submitted as we get follow up.
|
| |
|
Event or Problem Description
|
|
Sf 02677497 re charger- hot and melted.It was reported- hcp called tech support and reported that pt's charger got hot and melted at the junction of the charging cord and charger.Issue occurred on june 6th.Reportedly a new charger has already been sent to pt, but hcp was advised to call to report the issue.He said the charger with the issue has not been received yet, but he will send it to us.Hcp declined harm to pt.He read the notes from another provider that indicated the charger is still functioning, but damaged and occurred 1 time.Pt thinks a possible power surge.I advised to stop using the damaged charger.Hcp is unsure if has were in charger when the issue occurred.If has were in charger when the issue occurred, i advised to send the has in for evaluation as well.Qe form emailed to hcp.Flagged in navision.Complaint created.No harm was made to the user.This is a final report.No further follow up is expected.
|
| |
|
Additional Manufacturer Narrative
|
|
The actual device has passed both pcba test and final device test.Root cause analysis done in trended case states- by visual inspection it is observed that the surroundings of the usb interface has been deformed by heating.The usb connector in the device, as well as the connector on the cable are further investigated with a microscope.Functional testing reveals that the charger works fine, when running on the internal battery.That means that the malfunction in the device is purly a breakdown of the micro usb connector.The connector is broken, probably because the user has (not intentionally) misused it, while inserting the cable.If the cable has been slightly out of specification that could have had a negative effect on this.This has caused a low ohmic connection in the cable connector - almost a short circuit - that will emit heat when the cable is inserted into the wall charger, causing up to the rated 5 w of the wall charger to be dissipated inside the connector.Micro usb is a very well known and used standard but can mechanically break down.08sep2022: the case has been reviewed from a clinical perspective.Customer reported: the suspected device is a charger chgr premv3,c-1,gen,cff3.It was reported- while plugged in, the charger became hot and melted at the junction of the charging cord and the charger, the end user believes there may have been a power surge, it is unknown if the hearing aids were in the charger at the time of the incident and that the end user was not harmed and did not seek medical attention.Clinical evaluation of the event: from the previous investigations it is concluded that the high temperatures are generally limited to places difficult to touch, such as the back of the charger.Clinical conclusion on the case is that, even though unpleasant, it is evaluated unlikely that the device in failure mode would cause severe harm.R&d investigation results: investigation from this case revealed that the micro usb interface has broken down, causing the device to become hot.The charger and cable shows visible signs of heating.The area around the connector is deformed, as well as the cable connector.A substance is seen inside the connector and also below the connector on backside housing.The components on pcb around the connector seems fine.Micro usb is a very well known and used standard but can mechanically break down.Additional information: a dhr review have been completed.No deviation or changes were found during manufacturing of the device.Clinical evaluation performed.New code added for investigation finding: 4309-cause traced to environment the risk register will be updated to include a new warning/caution.This is a final report.No further information is expected.
|
| |
|