| Catalog Number |
982112D |
| Medical Device Problem Codes |
Peeled/Delaminated (1454); Device Difficult to Setup or Prepare (1487); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed as no product sample was received for evaluation.If the product is returned, the manufacturer will reopen this complaint for further investigation.No corrective actions are required since the complaint was not confirmed.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
|
| |
|
Event or Problem Description
|
|
It was reported that threading occurred when connecting the vacutainer to the needle.There were two (2) accidental needle pokes due to the plastic band that held the tubing is difficult to remove.No patient injury was reported.
|
| |
|