• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET WITH INJECTION SITE 20G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 WINGED INFUSION SET WITH INJECTION SITE 20G X 0.75IN; NEEDLE, HYPODERMIC, SINGLE LUMEN Back to Search Results
Model Number N/A
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/15/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported via chat "all 5 grippers that i have used so far leaks from a crack on the plastic tubing.We received a box of grippers a week ago and we have used five of the so far.I receive several different medication via pumps and syringes (push).After a few uses, a crack develops and medication leaks out.The crack is located in the clear tubing between the y fitting and the needle.These devises were purchase at a local home care service provider in canada.So far each and every one of them has failed after 1 day of use.¿ add info rec'd 08/24/2022: it was reported by ms&s " contacted customer over the phone.We discussed his situation ¿ he has a daughter that has a dual lumen ivad that requires dual access.When she had a single lumen ivad they were using gripper needles from smiths medical (and he said that this is what the lihn provided him with) and they worked great, but when she went to a dual lumen, the grippers didn¿t fit side by side.He stated that the hospital inserted our bard¿ wis 20g with y-site 0.75" for her dual lumen access so the care provider asked the lihn for these but they didn¿t provide them.So he sourced out a place to buy ours instead (cardinal) so that she could have dual access (she uses one lumen for tpn only and the other lumen for ns and meds).Then he noticed that after 2-3 days of infusion the administration set starts to leak.He says that most often it is nurses inserting the access needles and other times it is him, but they have done this for a long time and have not changed anything.I could not think of a reason for this leakage from the tubing based on the information provided.From the information gathered, it seems like this needle set is being used for an appropriate purpose.It was reported this occurred with five devices.This report addresses the fifth device.
 
Additional Manufacturer Narrative
The date of event was not provided by the complainant/reporter, the date reflected in this report is the date bd became aware of the event.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WINGED INFUSION SET WITH INJECTION SITE 20G X 0.75IN
Common Device Name
NEEDLE, HYPODERMIC, SINGLE LUMEN
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
becky garcia
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15328091
Report Number3006260740-2022-03405
Device Sequence Number17477750
Product Code FMI
UDI-Device Identifier00801741065866
UDI-Public(01)00801741065866
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K863606
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number2205220
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/15/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-