| Model Number |
9004050 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there were high readings.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Dhr (device history review) review is not relevant because the results of the complaint investigation do not indicate a problem with the initial manufacture or prior repair of the device.The device was received for evaluation.Visual and functional testing were performed.Visual inspection found the tamper seal was intact with an incorrect rear filter attached.Functional testing found confirmed the reported complaint.The device was reading out of specification.The root cause of the reported issue was found to be co2 bench was out of calibration; the device was calibrated.No product information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
|
| |
|