| Catalog Number |
03P88-25 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Apoc incident # (b)(4).Note: expiration date for cg8+, lot# w21351= 08/16/2022 (not exporting to submission report).Apoc labeling will be evaluated during the investigation as pertaining to the event.
|
| |
|
Event or Problem Description
|
|
On 08-aug-2021, abbott point of care was contacted by a customer regarding i-stat cg8+ cartridges that yielded suspected discrepant pco2 result of 120 mmhg on a 76 day old male patient.There was limited information and no patient information.Return product is not available for investigation.Method: i-stat, date: (b)(6) 2022, tested: 06:44, pco2(mmhg) 120; abl 90, (b)(6) 2022, 06:48, 63.There are no injuries associated with this event.At this time there is no reason to suspect a malfunction exists.The reporting decision was based on limited information available that suggests the product was not performing within the variability of the assay.Product lot informtaion was not presented, but requestged.The investigation is underway.
|
| |
|
Additional Manufacturer Narrative
|
|
Apoc incident: (b)(4).The investigation was completed on 05-oct-2022.A review of the device history record confirmed the cartridge lot met finished goods release criteria.Due to the expiration of cg8+ cartridge lot w21351, retained cartridge testing could not be performed.Pco2 performance verification shows that the overall performance of the i-stat is in agreement with other manufacturers.No deficiency has been determined for cg8+ cartridge lot w21351.
|
| |
|
Search Alerts/Recalls
|