Model Number 10000780
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/09/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for evaluation and the complaint was confirmed.No radiographs or images provided but it was reported to be flush in the set screw.Examination of the returned device identified excessive force as the likely cause of the tip fracture however, review of the manufacturing history found the device was release in the field over 6 years ago suggesting the root cause to be the result of age wear fatigue and excessive forces.The patient experienced no adverse consequence as a result of the failure.No additional investigation required.Label review: ".Do not implant the instruments: complications to the patient may include, but are not limited to: breakage of the device, which could make necessary removal difficult or sometimes impossible, with possible consequences of late infection and migration." ".Pre-operative warnings: the instruments should be treated as any precision instrument and should be carefully placed on trays, cleaned after each use, and stored, according to generally accepted hospital methods and practices.The instruments should be carefully examined prior to use for functionality, excessive wear, or damage.A damaged instrument should not be used as this may increase the risk of malfunction and potential patient injury." ".Intra-operative warnings: if any broken fragments of instruments remain in the body of a patient, they could cause allergic reactions or infections.If an instrument breaks in surgery and fragments go into the patient, these pieces should be removed prior to closure and should not be implanted." (b)(4).
 
Event or Problem Description
On (b)(6) 2022 a patient underwent a posterior fixation procedure.During final tightening the driver tip fractured off in the fenestrated screw at left l4.It was reported to be cold welded into the engaged screw and was left in patient encompassed flush in the set screw.No adverse consequences to the patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE GENERAL INSTRUMENTS
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENTS
Manufacturer (Section D)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INCORPORATED
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key15329133
Report Number2031966-2022-00168
Device Sequence Number20155444
Product Code LHX
UDI-Device Identifier00887517561039
UDI-Public887517561039
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number10000780
Device Lot NumberNM2627
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/11/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/19/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1