BIOMET 3I OSSEOTITE® CERTAIN® 2 IMPLANT 5 X 10MM; DENTAL IMPLANT Back to Search Results
Model Number XIFOSS510
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/11/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Zimvie complaint (b)(4).Patient weight is not provided / unknown.Initial reporter¿s title and last name are not provided / unknown.
 
Event or Problem Description
It was reported that while placing the implant, the threads was stripped and the implant overturned.Tooth # 30.The site was grafted prior to original implant placement.
 
Event or Problem Description
No additional or corrected information to report.
 
Additional Manufacturer Narrative
This report is being submitted to report additional information.The following sections are being reported: d4: expiration date and unique identifier (udi) number; h2: if follow-up, what type; h3: device evaluated by manufacturer; h4: device manufacturer date; h6: type of investigation; h6: investigation findings; h6: investigation conclusions; h10: additional narratives/data.The implant was returned, and the reported event (damaged threads) was confirmed.Based on the evaluation, device malfunction has occurred for the implant.There is no existing nonconformance / capa / hhe/d / ie / product hold against the reported devices that did or could cause or contribute to the reported event.Monthly post market trending review identified no actionable trends or corrective actions for the reported events or devices.Zimvie quality management system (qms) has controls in place to ensure the distribution of conforming products.Therefore, based on the available information, the products were within specifications and conforming when they left zimvie.Dhr review was completed for the subject lot number.It was confirmed that all operations and inspections were executed as per applicable procedures.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.A definitive root cause could not be determined.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OSSEOTITE® CERTAIN® 2 IMPLANT 5 X 10MM
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15329265
Report Number0001038806-2022-01344
Device Sequence Number6362188
Product Code DZE
UDI-Device Identifier00844868023210
UDI-Public(01)00844868023210(17)230508(10)2018050048
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K100724
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2023
Device Model NumberXIFOSS510
Device Catalogue NumberXIFOSS510
Device Lot Number2018050048
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/23/2022
Supplement Date Received by Manufacturer01/28/2023
Initial Report FDA Received Date08/31/2022
Supplement Report FDA Received Date02/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/08/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexMale
-
-