| Model Number |
00884450011676 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The account alleges that during an elective procedure for peripheral arterial occlusion between right femoral and popliteal arteries the wire broke while removing.Although the vessels had calcification, no significant tortuosity has not been found.Contralateral approach from left femoral artery was used for the procedure.After the completion of the infusion with the influsion catheter, the occluding wire was attempted to be withdrawn, leaving the infusion catheter in the vessel.At this time, the operator felt physical resistance at the vicinity of a y-connector at the operator's [proximal] end of the product.However, the operator pulled the wire out, and the wire got broken off.The breakage site was more proximal to the ball at the tip of the wire.The entire portion of the wire was withdrawn out with theinfusion catheter, deeming that the ball with the broken portion of the wire would not transfer into patient's body from the catheter tip.After the withdrawal, it was found that the catheter had kinked.The kink was not found before use.The planned procedure was completed.No patient injury to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was returned for evaluation.The complaint was confirmed and the root cause is attributed to clinical use and excessive force during use.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|
Search Alerts/Recalls
|