U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION RFA, LUER LOCK - LUER LOCK; SYRINGE, PISTON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION RFA, LUER LOCK - LUER LOCK; SYRINGE, PISTON Back to Search Results
Catalog Number H93813901
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reporter stated that the event occurred ¿over the last week¿.Initial reporter address: (b)(4).The actual devices were not available; however photos of actual samples were received for evaluation (unknown how many were for this report; lot was not documented in the photos).Visual inspection of the photographs identified white particulate matter on the devices.The reported condition was verified on the actual samples.The cause of the condition could not be determined.Additionally, five (5) unopened companion samples were received for evaluation.Visual inspection along with magnification was performed which did not identified any abnormalities that could have contributed to the reported condition; no particulate matter was identified.The reported condition was not verified on the companion samples.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that white specs were observed on an unspecified quantity of rapidfill connectors (red).This was observed when opening the packaging, prior to use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFA, LUER LOCK - LUER LOCK
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15331589
Report Number1416980-2022-04550
Device Sequence Number20452597
Product Code FMF
UDI-Device Identifier00085412478364
UDI-Public(01)00085412478364
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K872743
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date03/31/2025
Device Catalogue NumberH93813901
Device Lot Number60361276
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-