Catalog Number 1272.ED27
Device Problem Material Fragmentation (1261)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 08/15/2022
Event Type  malfunction  
Event Description
We have been informed that during posterior surgery, when removing the cannula, the metal part broke and fell outside the operative area.There is no report of patient/user harm.No report of prolonged or delayed surgery.
 
Manufacturer Narrative
The complaint is under investigation.
 
Event Description
We have been informed that during posterior surgery, when removing the cannula, the metal part broke and fell outside the operative area.There is no report of patient/user harm.No report of prolonged or delayed surgery.
 
Manufacturer Narrative
In regard to this complaint, one broken cannula was received.The capillary that is situated beneath the yellow ring had broken off.The funnel above the ring appeared to be missing.Close examination showed deformation of the capillary at the breaking point, suggesting that the capillary itself was grabbed with pliers or forceps, rather than the yellow ring.Though the observed deformation may have caused the break, it might just have occurred when the broken capillary was removed from the eye after breaking.Device history record review revealed no deviations and a database search showed that no similar complaints have been reported on any lot previously.Based on the investigation performed, the cause of failure could not be determined.The risk identified is included in the risk management documentation.Trend analysis indicates that the product is performing within anticipated rates.Complaints will be closely monitored to identify any significant adverse trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
27G CANNULA SET WITH HIGH FLOW INFUSION LINE
Type of Device
DORC TROCAR SYSTEM,
Manufacturer (Section D)
D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT.
scheijdelveweg 2
zuidland, 3214 VN
NL  3214 VN
Manufacturer (Section G)
D.O.R.C. DUTCH OPHTHALMIC RESEARCH CENTER INT.
scheijdelveweg 2
zuidland, 3214 VN
NL   3214 VN
Manufacturer Contact
danielle sleegers
scheijdelveweg 2
zuidland, 3214 -VN
NL   3214 VN
MDR Report Key15331689
MDR Text Key305363142
Report Number1222074-2022-00059
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1272.ED27
Device Lot Number2000431226
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-