Catalog Number 367835
Medical Device Problem Code Short Fill (1575)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4mg blood collection tubes, the device experienced underfill or low draw of a tube with blood.This event occurred 60 times.The following information was provided by the initial reporter.The customer stated: tubes are presenting vacuum failure.The tube fills only 1 to 2 ml, when it should has filled 3ml, and does not accomplishes with additive/blood ratio.How many units of the product presented the reported deviation? (amount affected);of each box of 100, 60 were faulty.
 
Additional Manufacturer Narrative
A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H.6.Investigation summary: bd had not received samples or photos for investigation.Therefore, 10 retention samples from bd inventory were evaluated by functional testing and no issues were observed relating to underfill as all samples met specifications.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.There were no related quality issues during manufacturing of the product.This complaint is unable to be confirmed for the indicated failure mode underfill.Bd was not able to identify a root cause for the indicated failure mode.The quantity of blood drawn into evaluated tubes varies with altitude, ambient temperature, barometric pressure, tube age, venous pressure, and filling technique.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported when using the bd vacutainer® k2 edta (k2e) 5.4mg blood collection tubes, the device experienced underfill or low draw of a tube with blood.This event occurred 60 times.The following information was provided by the initial reporter.The customer stated: tubes are presenting vacuum failure.The tube fills only 1 to 2 ml, when it should has filled 3ml, and does not accomplishes with additive/blood ratio.¿ how many units of the product presented the reported deviation? (amount affected); a: of each box of 100, 60 were faulty.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® K2 EDTA (K2E) 5.4MG BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer (Section G)
BECTON, DICKINSON & CO. (BROKEN BOW)
150 south 1st avenue
broken bow NE 68822
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15331874
Report Number1917413-2022-00528
Device Sequence Number6478883
Product Code JKA
UDI-Device Identifier00382903678358
UDI-Public00382903678358
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
BK050036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Catalogue Number367835
Device Lot Number1195777
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/09/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date08/31/2022
Supplement Report FDA Received Date09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-