Model Number EPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Medical Device Problem Code Unexpected Shutdown (4019)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported their epiq cvxi ultrasound system shut down unexpectedly during a cardiac mitral valve procedure.The procedure in progress was completed successfully using a different ultrasound system available at the customer site.There was no harm or altered patient outcome as a result of the issue.
 
Additional Manufacturer Narrative
Initial evaluation of the system logs indicate defective batteries to be the cause of this issue.Repair details and final system logs evaluation will be included in a follow up report upon system repair and investigation completion.
 
Additional Manufacturer Narrative
In response to the issue, a service escalation was initiated.Philips technical support performed a thorough review of system log files and found several entries indicating defective batteries.As a correction, the batteries were replaced.The suspect batteries were disposed of locally so no further failure analysis was realized.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIQ CVXI
Common Device Name
SYSTEM, IMAGING, PULSED DOPPLER, ULTRASONIC
Manufacturer (Section D)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer (Section G)
PHILIPS ULTRASOUND
22100 bothell everett highway
bothell WA 98021
Manufacturer Contact
eric nail
22100 bothell everett highway
bothell, WA 98021
MDR Report Key15331940
Report Number3019216-2022-00081
Device Sequence Number15187826
Product Code IYN
UDI-Device Identifier00884838097940
UDI-Public00884838097940
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K132304
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEPIQ DIAGNOSTIC ULTRASOUND SYSTEM
Device Catalogue Number795232
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date08/31/2022
Supplement Report FDA Received Date10/13/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/21/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-