Model Number 500-301
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Great Vessel Perforation (2152); Pericardial Effusion (3271)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device lot number, expiration date unk.The device was discarded, thus no investigation could be completed.Device manufacture date unk because lot number unk.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
 
Event or Problem Description
A lead extraction procedure commenced to remove a right atrial (ra) and a right ventricular (rv) lead due to cied system/pocket infection.Spectranetics lead locking devices were inserted into each lead to provide traction.Using a spectranetics 12f glidelight laser sheath and a spectranetics visisheath dilator sheath, the ra lead was successfully removed.The patient's blood pressure dropped rapidly, and a pericardial effusion was detected by transesophageal echocardiography (tee).The patient's blood pressure dropped to zero caused by cardiac tamponade.A sternotomy was immediately performed and a 1.5 cm superior vena cava (svc) perforation was discovered.A rescue balloon was inflated to control bleeding, and the patient was placed on bypass.The repair of the svc perforation was successful.The rv lead was also removed during the procedure, and the patient survived.This report captures the 12f glidelight in use when the svc perforation occurred, requiring intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECTRANETICS 12F GLIDELIGHT LASER SHEATH
Common Device Name
DEVICE, REMOVAL, PACEMAKER ELECTRODE, PERCUTANEOUS
Manufacturer (Section D)
THE SPECTRANETICS CORPORATION
9965 federal drive
colorado springs CO 80921
Manufacturer (Section G)
SPECTRANETICS
9965 federal drive
colorado springs CO 80921
Manufacturer Contact
barbara creel
9965 federal drive
colorado springs, CO 80921
MDR Report Key15332051
Report Number1721279-2022-00165
Device Sequence Number6485981
Product Code MFA
UDI-Device Identifier00813132020323
UDI-Public00813132020323
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P960042 S070
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number500-301
Device Catalogue Number500-301
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/10/2022
Initial Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BOSTON SCIENTIFIC 0185 RV PACING LEAD.; BOSTON SCIENTIFIC 4473 RA PACING LEAD.; SPECTRANETICS CVX-300 EXCIMER LASER SYSTEM.; SPECTRANETICS LEAD LOCKING DEVICES.; SPECTRANETICS VISISHEATH DILATOR SHEATH.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age48 YR
Patient SexFemale
Patient Weight45 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-