MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYES 10MM 30°, HD AUTOCLAVABLE LAPAROSCOPE 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PRECISION IDEAL EYES 10MM 30°, HD AUTOCLAVABLE LAPAROSCOPE 33CM; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number 0502103030
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/18/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was foggy image.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
This scope was not received for evaluation at stryker endoscopy.This scope was received at henke for repair, but was not evaluated therefore the reported failure "constant fogging of the optics" was not confirmed.According to henke: outer tube damaged, distal tip, distal tip damaged, - outer tube body / light post, body damaged.Probable root cause: laser welding seal failure, distal/proximal window solder failure, damage to optical train, damage to needle, damage to distal or proximal windows, moisture intrusion, end of life wear-out, environmental disturbance: endoscope colder than dew point, environmental disturbance: fluid entrapment between the coupler and eyepiece, junction (eyepiece only), use error.The device manufacture date is not known the reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION IDEAL EYES 10MM 30°, HD AUTOCLAVABLE LAPAROSCOPE 33CM
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
chester rebugio
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15332283
Report Number0002936485-2022-00499
Device Sequence Number20442787
Product Code GCJ
UDI-Device Identifier07613327062182
UDI-Public07613327062182
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K910132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial,Followup
Report Date (Section B) 10/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502103030
Device Catalogue Number0502103030
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/19/2022
Supplement Date Received by Manufacturer08/19/2022
Initial Report FDA Received Date08/31/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-