| Model Number |
4585 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Per medwatch received: "battery compartment on wrist device forcefully separated while in use on patient." no consequences or impact to patient were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional manufacturing narrative: the returned device was evaluated.Visual inspection found the sensor housing has separated.Battery voltage testing unavailable because the white portion of sensor housing (including batteries) was not received for investigation.No short or open was detected, and no intermittent short or open was detected when sensor is manipulated.Unable to test device functionality due to damaged sensor.
|
| |
|