• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ONKOS SURGICAL ELEOS; STEM EXTENSION, CEMENTED, 14MMX65MM, STRAIGHT, FLUTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ONKOS SURGICAL ELEOS; STEM EXTENSION, CEMENTED, 14MMX65MM, STRAIGHT, FLUTED Back to Search Results
Model Number KSC01465E
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Insufficient Information (4580)
Date of Event 08/01/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is in progress.When it is complete a supplemental mdr will be submitted accordingly.Multiple mdrs were submitted for this adverse event: 3013450937-2022-00245, 3013450937-2022-00246, 3013450937-2022-00247, 3013450937-2022-00248, 3013450937-2022-00249, 3013450937-2022-00250, 3013450937-2022-00252 and 3013450937-2022-00253.
 
Event or Problem Description
It was reported that a patient experienced an infection and underwent a revision surgery on (b)(6) 2022.The following implants were removed during the revision: eleos distal femur axial pin, eleos tibial hinge w/ rotational stop and eleos tibial poly spacer.
 
Additional Manufacturer Narrative
This report is being submitted to include additional information.The investigation is complete.The device was not returned for evaluation.The root cause of the alleged infection could not be determined.The device history records and sterilization batch records were reviewed and no issues during manufacturing or sterilization were identified that would have contributed to this complaint.If any additional information is obtained, a supplemental mdr will be filed accordingly.The following sections were updated: h3: device evaluated by manufacturer updated to no.H6: type of investigation code updated to 3331: analysis of production records.H6: type of investigation code updated to 4111: communication/interviews.H6: type of investigation code updated to 4110: trend analysis.H6: type of investigation code updated to 4114: device not returned.H6: type of investigation code updated to 4117: device not accessible for testing.H6: type of investigation code updated to 4109: historical data analysis.H6: investigation findings code updated to 213: no device problem found.H6: investigation conclusions code updated to 4315: cause not established.
 
Event or Problem Description
It was reported that a patient developed an alleged infection and underwent a revision surgery on (b)(6) 2022.The following implants were removed during the revision: eleos distal femur axial pin, eleos tibial hinge w/ rotational stop and eleos tibial poly spacer.The following implants remain implanted from this patient's original surgery: eleos distal femur, two eleos tibial block augments, eleos cemented stem extension, eleos segmental stem, and eleos tibial baseplate.No additional information regarding this adverse event has been reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEOS
Common Device Name
STEM EXTENSION, CEMENTED, 14MMX65MM, STRAIGHT, FLUTED
Manufacturer (Section D)
ONKOS SURGICAL
77 east halsey road
parsippany NJ 07054
Manufacturer (Section G)
MICROPORT ORTHOPEDICS INC
5677 airline rd
arlington TN 38002
Manufacturer Contact
danielle pierce
77 east halsey road
parsippany 07054
8447672766
MDR Report Key15332466
Report Number3013450937-2022-00251
Device Sequence Number6937692
Product Code KRO
UDI-Device IdentifierB278KSC01465E0
UDI-PublicB278KSC01465E0
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K161520
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberKSC01465E
Device Catalogue NumberKSC01465E
Device Lot Number1866005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/01/2022
Supplement Date Received by Manufacturer10/28/2022
Initial Report FDA Received Date08/31/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/13/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
25000007E ELEOS DISTAL FEMUR, LEFT, SEGMENTAL; 25001208E ELEOS TIBIAL POLY SPACER 8MM; 25002111E ELEOS DISTAL FEMUR AXIAL PIN; CS-15120-03M ELEOS CEMENTED SEGMENTAL STEM; KTAGB305E ELEOS TIBIAL BLOCK AUGMENT SIZE: 3; KTAGB305E ELEOS TIBIAL BLOCK AUGMENT SIZE: 3; TB-2203E-01M ELEOS TIBIAL BASEPLATE SIZE: 3; THSMWRS01M ELEOS TIBIAL HINGE W/ROTATIONAL STOP
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age53 YR
Patient SexFemale
-
-