• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA LINER: VERSAFITCUP DM DOUBLE MOBILITY HC LINER Ø 50/28; HIP LINER DOUBLE MOBILITY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA LINER: VERSAFITCUP DM DOUBLE MOBILITY HC LINER Ø 50/28; HIP LINER DOUBLE MOBILITY Back to Search Results
Model Number 01.26.2850MHC
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 08/04/2022
Type of Reportable Event Serious Injury
Event or Problem Description
At about 1 month after the primary surgery, the patient came in due to signs of an infection and the pathogen is unknown.The surgeon performed a washout and revised the head and liner.The surgery was completed successfully.
 
Additional Manufacturer Narrative
Batch review performed on (b)(6) 2022.Lot 2118106: 110 items manufactured and released on 01-feb-2022.Expiration date: 2027-01-19.No anomalies found related to the problem.To date, 71 items of the same lot have been sold with no similar reported cases during the period of review.Additional component involved: ball heads: mectacer 01.29.201 biolox delta ceramic ball head 12/14 ø 28 size s -3.5 (k112115) lot.2114357: 150 items manufactured and released on 10-jan-2022.Expiration date: 2026-12-21.No anomalies found related to the problem.To date, 105 items of the same lot have been sold with no similar reported cases during the period of review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINER: VERSAFITCUP DM DOUBLE MOBILITY HC LINER Ø 50/28
Common Device Name
HIP LINER DOUBLE MOBILITY
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15333461
Report Number3005180920-2022-00664
Device Sequence Number20588235
Product Code MEH
UDI-Device Identifier07630030807282
UDI-Public07630030807282
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K092265
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number01.26.2850MHC
Device Catalogue Number01.26.2850MHC
Device Lot Number2118106
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2022
Initial Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-