| Model Number |
8401 |
| Medical Device Problem Codes |
Device Alarm System (1012); Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found no damage.Functional testing found occlusion error was present after powering up the monitor.Opened the monitor and found the tubing going to the bench was kinked.The root cause of the reported issue was found to be a known design issue where the weight of the restrictor pushes the small diameter tubing against the co2 bench nozzle as the housing closes.Tubing kit was replaced and monitor passed all functional testing.Operator of device is unknown.Initial reporter also sent report to fda is unknown.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the monitor reads continuous occlusion and was unable to obtain end tidal reading.No patient injury was reported.
|
| |
|