ST PAUL BCI CAPNOCHECK II, MODEL 8400; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 8401
Medical Device Problem Codes Device Alarm System (1012); Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection found no damage.Functional testing found occlusion error was present after powering up the monitor.Opened the monitor and found the tubing going to the bench was kinked.The root cause of the reported issue was found to be a known design issue where the weight of the restrictor pushes the small diameter tubing against the co2 bench nozzle as the housing closes.Tubing kit was replaced and monitor passed all functional testing.Operator of device is unknown.Initial reporter also sent report to fda is unknown.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# (b)(4).
 
Event or Problem Description
It was reported that the monitor reads continuous occlusion and was unable to obtain end tidal reading.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI CAPNOCHECK II, MODEL 8400
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
6430 post rd
minneapolis, MN 55442
MDR Report Key15334166
Report Number3012307300-2022-17156
Device Sequence Number17761942
Product Code CCK
UDI-Device Identifier10610586036781
UDI-Public10610586036781
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8401
Device Catalogue Number8401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Initial Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-