| Model Number |
LCB033 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/19/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No procedure delays have been reported.Retain testing was performed on the lot number subject of the reported event; no issues were noted.The dhr for the subject lot was reviewed and no abnormalities were found.The process challenge device instructions for use (pg.1) states, "instructions for use: before use, examine the pcd to ensure that the foil and seal are intact.Examine the integrator through the plastic, to ensure the pcd has not been used previously." no additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
The user facility reported via medwatch (mw5111090) that they received "dried up" control containers and that their biological tests have been found dry prior to and after processing.No report of injury.
|
| |
|