| Model Number |
TMTB10 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the implant fracture during placement.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimmer biomet complaint number (b)(4).
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
An imp tm 3.7mm mtx full,10m (tmtb10) was returned for investigation.Visual inspection of the as returned product identified worn marking due to usage, bone and a fracture at the bottom of the implant.No pre-existing conditions were noted on the per.The reported device was location is #19 and was used same day.Pictures or x-ray images were not provided.Appropriate documentation was reviewed.Dhr review was completed for the subject lot number (1244887).It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformances, which could have caused or contributed to the reported event was noted as part of the dhr.Lot was inspected and passed all acceptance criteria by qa.Complaint history review was performed for the reported lot number (1244887) for similar event and no other complaint was identified.Nov post market trending was reviewed and there were no actionable events or corrective actions for the reported event or device.Based on the available information, device malfunction did occur, and the reported event was confirmed.
|
| |
|