| Catalog Number |
OSC-5F-10T/B |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The suspect device has returned for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges duriing a drainage procedure, the drainage catheter valves were leaking fluid as the needle was being removed.No patient injury to report.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device has been returned for evaluation.The complaint is confirmed.The root cause is attributed to the manufacturing process.A search of the complaint database was performed and no similar complaints for this lot number were found.The device history record was reviewed, and no exception documents were found.
|
| |
|