BTG YOKNEAM ICESPHERE 1.5 CX 90 DEGREE; UNIT, CRYOSURGICAL, ACCESSORIES Back to Search Results
Model Number FPRPR3573
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2022
Type of Reportable Event Malfunction
Event or Problem Description
A shaft gas leak was reported.A 1.5 cx 90 degree icephere needle was selected for use in a renal cryoablation procedure.During device integrity testing, prior to insertion into the patient, the needle was submerged in a testing bowl and bubbles occurred along the shaft of the needle at the proximal end near the shaft to handle joint, but not at the joint.The needle was exchanged and the procedure was completed with no reported patient complications.
 
Event or Problem Description
A shaft gas leak was reported.A 1.5 cx 90 degree icephere needle was selected for use in a renal cryoablation procedure.During device integrity testing, prior to insertion into the patient, the needle was submerged in a testing bowl and bubbles occurred along the shaft of the needle at the proximal end near the shaft to handle joint, but not at the joint.The needle was exchanged and the procedure was completed with no reported patient complications.
 
Additional Manufacturer Narrative
Device evaluated by mfr: the complaint device was returned and tested for an air leak with none found.No air bubbles were seen coming from the needle shaft or the handle joints.The device was ran through a freeze cycle and produced an iceball of adequate size.There was no issues found with the functionality of the device and the complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICESPHERE 1.5 CX 90 DEGREE
Common Device Name
UNIT, CRYOSURGICAL, ACCESSORIES
Manufacturer (Section D)
BTG YOKNEAM
tavor bldg 1, industrial park
po box 224
yokneam 20692 03
IS  2069203
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
tavor bldg number 1
yokneam 20692 03
IS   2069203
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15335922
Report Number2134265-2022-08709
Device Sequence Number6478081
Product Code GEH
UDI-Device Identifier07290104831133
UDI-Public07290104831133
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/18/2024
Device Model NumberFPRPR3573
Device Catalogue NumberFPRPR3573
Device Lot Number0029132562
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/08/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date09/01/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/29/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexFemale
-
-