Model Number 8401
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history review (dhr) was not applicable or relevant because the results of the complaint investigation did not indicate a problem with the initial manufacture and this was the first time the unit came in for service.The product was received for evaluation.Visual and functional testing were performed.A visual inspection found the device to be in excellent physical condition.Functional testing by powering on the returned monitor found that when the monitor was manually shook an occlusion error displayed.Internal inspection did not find anything that would have caused the occlusion error; however, the absorber was known to cause occlusion error due to small white particles on the inside of the carbon dioxide (co2) absorber which could have clogged the air flow path and caused the monitor to occlude.The root cause of the reported issue was found to be that the absorber caused the occlusions.The technician replaced the absorber.No information has been provided to date.This remediation mdr was generated under protocol (b)(4), as a result of warning letter (b)(4).
 
Event or Problem Description
It was reported that there was an occlusion error.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPNOCHECK II CAPNOGRAPH SYSTEM
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15336093
Report Number3012307300-2022-17265
Device Sequence Number12494430
Product Code CCK
UDI-Device Identifier10610586036781
UDI-Public10610586036781
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8401
Device Catalogue Number8401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2022
Initial Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/20/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-