| Catalog Number |
339.31 |
| Medical Device Problem Code |
Insufficient Heating (1287)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that an ophthalmic diathermy probe exhibited insufficient heat before surgery.The surgery was performed and completed after replacing the product with another one.There was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
A product sample has been received by the manufacturer and it is awaiting evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
No lot number was identified with this complaint; therefore, the manufacturing documentation was not reviewed.All product and batch history records are quality reviewed prior to product release.The instrument was visually inspected.No abnormality was observed.Electrical resistance measurements indicate insufficient contact.When the instrument was opened, the contacts were not destroyed.Nevertheless, no defect contact points in the instrument could be identified.All contact points were specifically investigated but none of the contact points was loose.Based on the outcome of this investigation the complaint of the customer can be confirmed but the root cause could not be determined anymore.In production a 100% inspection of the instrument is performed which would identify defect contact points.Since the instrument passed this test, the contact must have been affected later.But this cannot be determined anymore.The root cause cannot be identified conclusively because the sample has pass the 100% inspection.As the blister was opened by the customer, he handled the product.The point in time when the malfunction occurred, can no longer be determent.This complaint has been reviewed and future data will be monitored for evidence of adverse trending and further action will be taken.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|