• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC 11.0 CM LONG ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC 11.0 CM LONG ATTACHMENT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) Back to Search Results
Model Number LONG
Medical Device Problem Code Excessive Heating (4030)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported from taiwan that the attachment device had become unusually warm.It was not reported if the event occurred during a surgical procedure.It was not reported if there was a delay to a planned procedure.It was not reported if there was a spare device available for use.There was no patient involvement reported.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2022.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.Additional narrative: reporter's phone number was not provided.As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.Udi ¿ (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The attachment device was evaluated and the reported condition that the device had become unusually warm was not confirmed.Therefore, an assignable root cause was not determined.However, during evaluation it was observed that the nose tube was bent/damaged, the bearing was damaged, and the cutter device could not be inserted.It was further determined that the device failed pretest for visual assessment and cutter insertion assessment.The assignable root cause of these conditions was determined to be traced to component failure due to wear.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11.0 CM LONG ATTACHMENT
Common Device Name
DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC)
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
kate karberg
4500 riverside drive
palm beach gardens, FL 33410
3035526892
MDR Report Key15336759
Report Number1045834-2022-01173
Device Sequence Number6485761
Product Code ERL
UDI-Device Identifier00845384003571
UDI-Public00845384003571
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K011444
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberLONG
Device Catalogue NumberLONG
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/26/2022
Supplement Date Received by Manufacturer09/15/2022
Initial Report FDA Received Date09/01/2022
Supplement Report FDA Received Date09/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/23/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-