• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ARTICUL/EZE BALL 32 +13 BLK; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ARTICUL/EZE BALL 32 +13 BLK; ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 1365-24-000
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 08/16/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision completed due hip pain and dysfunction, large fragments of poly (determined as "petals" of locking bore of constrained liner) and metallosis present, locking ring loose/disengaged, dislocation without difficulty.Original stem and cup from 2014 revision stable and left in-situ.Surgeon aware that skirted head should not be used with constrained liner and was cause of failure.Previous revision completed due to instability/dislocation.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : the device associated with this report was not returned for analysis.Review of the photographic and x-ray evidence was able to confirm the reported dislocation event.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
In the primary procedure, the zimmer biomet fitmore acetabular component was used with our femoral components.The affected side involved in this event was the left hip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICUL/EZE BALL 32 +13 BLK
Common Device Name
ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15336783
Report Number1818910-2022-17049
Device Sequence Number17763895
Product Code JDI
UDI-Device Identifier10603295033202
UDI-Public10603295033202
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K883460
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1365-24-000
Device Catalogue Number136524000
Device Lot NumberD14042567
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/17/2022
Supplement Date Received by Manufacturer11/14/2022
11/14/2022
Initial Report FDA Received Date09/01/2022
Supplement Report FDA Received Date11/14/2022
11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/13/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PINN LNR CON +4 10D 32IDX56OD; PINN MULTIHOLE W/GRIPTION 56MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-