SMITHS MEDICAL ASD, INC. SAF-T HOLDER DEVICES; TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COL
|
Back to Search Results |
|
| Model Number |
96004 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No sample was received to perform an investigation; however, two pictures were received.Both pictures showed a non-original equipment manufacturer (oem) adaptor with a broken off luer inside.Based on the pictures received, the reported allegation was confirmed, but the cause could not be determined.A device history record (dhr) review was performed, and no issues were noted during manufacture.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the device broke apart at the luer lock adapter section leaving a piece embedded in the needle-less clave on the intravenous (iv) pigtail.The patient's pigtail needed to be removed due to a continuous leak.No patient injury was reported.
|
| |
|