• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. (AF-ST. PAUL) DISPOSABLE RADIOFREQUENCY ELECTRODE, 10CM; PROBE, RADIOFREQUENCY LESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. (AF-ST. PAUL) DISPOSABLE RADIOFREQUENCY ELECTRODE, 10CM; PROBE, RADIOFREQUENCY LESION Back to Search Results
Model Number RFDEUK-10
Medical Device Problem Code Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/08/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Patient information cannot be provided due to personal data privacy legislation/policy.
 
Event or Problem Description
Related manufacturer report numbers: 2184149-2022-00197, 2184149-2022-00198, 2184149-2022-00199, 2184149-2022-00200, 2184149-2022-00201, 2184149-2022-00202.During the procedure, there was a missing impedance error message, and the generator stopped working after the sensory and motor testing.The procedure could not be completed.However after the procedure, during troubleshooting, the electrode was replaced and the generator was functioning properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISPOSABLE RADIOFREQUENCY ELECTRODE, 10CM
Common Device Name
PROBE, RADIOFREQUENCY LESION
Manufacturer (Section D)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer (Section G)
ST. JUDE MEDICAL, INC. (AF-ST. PAUL)
one st. jude medical drive
st. paul MN 55117
Manufacturer Contact
ronnie shalev
6901 preston road
plano, TX 75024
9723098000
MDR Report Key15337392
Report Number2184149-2022-00203
Device Sequence Number6889773
Product Code GXI
UDI-Device Identifier05415067028990
UDI-Public05415067028990
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K011387
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2024
Device Model NumberRFDEUK-10
Device Catalogue NumberRFDEUK-10
Device Lot Number8313476
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Initial Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-