The eli 280 is a 12-lead electrocardiograph device intended to be used to perform 12 lead resting ecgs.The acquired data can be reviewed, stored, printed, and/or transmitted to an emr system.The eli 280 resting electrocardiograph is indicated for use to acquire, analyze, display, and print electrocardiograms.Device is indicated for use to provide interpretation of the data for consideration by a physician.Device is indicated for use in a clinical setting, by a physician or by trained personnel who are acting on the orders of a licensed physician.It is not intended as a sole means of diagnosis.The interpretations of ecg offered by the device are only significant when used in conjunction with a physician over-read as well as consideration of all other relevant patient data.Device is indicated for use on adult and pediatric populations.The device is not intended to be used as a vital signs physiological monitor.The device was returned for repair where it failed the functional check and it was determined that the issue was with the mainboard; the drive gear screw was loose on the printer transmission.The hrc technician reset the screw and torqued to spec using loctite.The device was evaluated, and safety tested and was functioning as designed.Ecgs are routinely performed to diagnose cardiac problems and there are no risks associated with the test itself, however, there is a risk of associating ecg records with wrong patients leading to misdiagnosis and incorrect treatment.For this reason, hillrom determines this event reportable.
|