• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL BCI CAPNOCHECK PLUS SLEEP CARBON DIOXIDE DETECTOR; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL BCI CAPNOCHECK PLUS SLEEP CARBON DIOXIDE DETECTOR; ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE Back to Search Results
Model Number 8401
Medical Device Problem Code Device Alarm System (1012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A device history record (dhr) review was not conducted, based upon review of the information provided by the customer, as it does not indicate a problem with the initial manufacture or prior repair of the device.The product was received for evaluation.Visual and functional testing were performed.Visual inspection showed the battery door was missing.The reported problem was confirmed.The root cause of the reported issue was found to be the desiccated absorber was creating an obstruction.The technician replaced the desiccated absorber.The device passed final functional test.This remediation mdr was generated under protocol b10009406, as a result of warning letter cms# 617147.
 
Event or Problem Description
It was reported that the device alarmed occlusion constantly and no occlusion.No patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BCI CAPNOCHECK PLUS SLEEP CARBON DIOXIDE DETECTOR
Common Device Name
ANALYZER, GAS, CARBON-DIOXIDE, GASEOUS-PHASE
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
jim vegel
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key15337817
Report Number3012307300-2022-17335
Device Sequence Number6486771
Product Code CCK
UDI-Device Identifier10610586036781
UDI-Public10610586036781
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K991086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8401
Device Catalogue Number8401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2022
Initial Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/05/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-