(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, d9, g3, g6, h1, h2, h3, h6, h10 complaint sample was evaluated, and the reported event was confirmed.Visual evaluation of the returned product found a hair-like debris in the sealed sterile pouch.Device history record (dhr) was reviewed, and no discrepancies were found.The condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|