| Model Number |
N/A |
| Medical Device Problem Codes |
Degraded (1153); Fracture (1260); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/12/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the adjustment screw of the intramedullary femoral guide stripped and fractured when being disassembled after being washed through central sterile.There is no additional information available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up is being submitted to relay additional information.Updated: b4, b5, g3, g6, h1, h2, h3, h6, and h10.Visual examination of the returned product found it to exhibit signs of repeated use (nicked / gouged) and components were disassembled, threads are damaged on both components.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.The root cause of the reported issue is attributed to normal wear of the device as it was used over six years.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|