• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. ADJUSTABLE INTRAMEDULLARY FEMORAL GUIDE; GUIDE, SURGICAL, INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. ADJUSTABLE INTRAMEDULLARY FEMORAL GUIDE; GUIDE, SURGICAL, INSTRUMENT Back to Search Results
Model Number N/A
Medical Device Problem Codes Degraded (1153); Fracture (1260); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the adjustment screw of the intramedullary femoral guide stripped and fractured when being disassembled after being washed through central sterile.There is no additional information available.
 
Additional Manufacturer Narrative
(b)(4).The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, g3, g6, h1, h2, h3, h6, and h10.Visual examination of the returned product found it to exhibit signs of repeated use (nicked / gouged) and components were disassembled, threads are damaged on both components.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.The root cause of the reported issue is attributed to normal wear of the device as it was used over six years.If any further information is which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADJUSTABLE INTRAMEDULLARY FEMORAL GUIDE
Common Device Name
GUIDE, SURGICAL, INSTRUMENT
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15339679
Report Number0001822565-2022-02526
Device Sequence Number6438922
Product Code FZX
UDI-Device Identifier00889024212923
UDI-Public(01)00889024212923(10)63402773
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number00596704200
Device Lot Number63402773
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/12/2022
Supplement Date Received by Manufacturer11/14/2022
Initial Report FDA Received Date09/01/2022
Supplement Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient SexUnknown
-
-