| Model Number |
1961-92-022 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526); Unspecified Tissue Injury (4559); Insufficient Information (4580)
|
| Date of Event |
08/16/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Clinical adverse event received for spin out event is serious and is considered moderate event is possibly related to device and is possibly related to procedure.Date of implantation: (b)(6) 2022.Date of event (onset): (b)(6) 2022.(left knee) treatment: revision of tibial insert.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # pc- (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Added: b5, h6 health effect - clinical code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Medical records received.On (b)(6) 2022, the patient had a left total cemented knee arthroplasty to address end-stage arthrosis.Depuy components, including depuy patella and depuy cement were used during this procedure.(b)(6) 2022 medical records note the patient is status post left knee replacement with new onset pain and popping.The initial surgery was 5 months ago, then the patient was septic postoperatively and had pneumonia as well as c.Difficile.It was noted on physical examination that the patient has lots of valgus laxity.The impression is that the patient has status post left knee replacement with mcl injury and likely spinning out of poly.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.All available x-rays and photographs were reviewed and it was not possible to identify any rotation or any other device issue trough x-ray.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation mre was not performed.
|
| |
|