Model Number TF85-32-63-45
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Heart Block (4444)
Date of Event 07/14/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
There is no reasonable evidence to suggest that the baylis medical devices caused or contributed to the reported incident.However, as baylis medical devices were reported to be among the devices used in the procedure, baylis medical has decided to submit this report.Dhr review was completed for the lot in question and all the devices fulfilled all requirements prior to release.
 
Event or Problem Description
A case of heart block requiring intervention during a left atrial appendage closure (laac) procedure was reported where the baylis medical nrg transseptal kit was used for transseptal puncture.As reported, the right bundle was inadvertently knocked while maneuvering the transseptal system, causing the patient to go into heart block.Cardiopulmonary resuscitation and external pacing were performed.The patient was stabilized and the laac procedure was successfully completed.There is no reasonable evidence to suggest that the baylis medical devices caused or contributed to the reported incident.However, as baylis medical devices were reported to be among the devices used in the procedure, baylis medical has decided to submit this report.A separate problem report has been submitted for the nrg transseptal needle with the report number 3019751610-2022-00033.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TORFLEX TRANSSEPTAL GUIDING SHEATH
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
BAYLIS MEDICAL COMPANY INC.
5959 trans-canada highway
montreal, quebec H4T 1 A1
CA  H4T 1A1
Manufacturer (Section G)
BAYLIS MEDICAL COMPANY INC.
2775 matheson blvd. east
mississauga, ontario L4W 4 P7
CA   L4W 4P7
Manufacturer Contact
meghal khakhar
2775 matheson blvd. east
mississauga, ontario L4W 4-P7
CA   L4W 4P7
MDR Report Key15341239
Report Number3019751610-2022-00028
Device Sequence Number5231453
Product Code DYB
UDI-Device Identifier00685447005391
UDI-Public(01)00685447005391(17)240302(10)TFB010322
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K183655
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date (Section B) 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberTF85-32-63-45
Device Lot NumberTFFB010322
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/08/2022
Initial Report FDA Received Date09/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-