| Model Number |
130-03-738 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Joint Laxity (4526); Implant Pain (4561)
|
| Date of Event |
08/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Expiration date: 12/16/2024 (16-dec-2024).The reason for this revision surgery was reported as right knee pain and loosening.The previous surgery and the surgery detailed in this event occurred 2.5 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at hospital and not made available to djo surgical for examination.A review of the device history record (dhr) shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There was no non-conforming material report (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to right knee pain.X-ray showed the patella implant was loose and had disassociated from the bone.There were no findings during this evaluation that indicate the reported device was defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
|
| |
|
Event or Problem Description
|
|
Revision surgery - patient presented to surgeon with right knee pain.On x-ray it appeared that the patella implant was loose and had disassociated from the bone.In the revision surgery, it was confirmed that the patella had come loose, and appeared that the pegs on the patella implant had sheared off.Prior bone cement was on the implant but did not appear on the native patella.Surgeon speculated that possibly in the primary surgery bone cement had setup quickly and perhaps did not get strong adherence to the bone.Pre-revision x-rays and intra-operative pictures attached.
|
| |
|
Search Alerts/Recalls
|