• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENDOLOGIX AFX2; BIFURCATED STENT GRAFT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENDOLOGIX AFX2; BIFURCATED STENT GRAFT Back to Search Results
Model Number BEA22-90/I16-30
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Ischemia (1942); Necrosis (1971); Multiple Organ Dysfunction Syndrome (3261); Embolism/Embolus (4438)
Date of Event 08/12/2022
Type of Reportable Event Death
Event or Problem Description
The patient was initially implanted with a bifurcated stent graft and an infrarenal aortic extension to treat an abdominal aortic aneurysm (aaa).It was reported that the patient had poor anatomy, described as severe calcification and a narrowed external iliac artery (eia).After the stent grafts were implanted the procedure was completed without any endoleaks.Four (4) days post implant a shower embolization event occurred and the patient deteriorated to a fatal condition.Per the physician, the internal iliac arteries (iia) were also found embolized.The patient's blood pressure changed and necrosis was observed in the lower legs at the time of the shower embolization.Dialysis was performed at that time.It was reported that the patient expired four days post implant.The cause of death is multi organ failure due to intestinal ischemia.Both the iia and the inferior mesenteric artery (ima) were occluded at the same time when the stent grafts were implanted and because of thrombi which could have lead to the ischemia event.
 
Additional Manufacturer Narrative
The device involved in this event will not be returned for evaluation as it remains implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by the clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.Device remains implanted.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good-faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the shower embolization is confirmed.The patient death is unconfirmed due to a lack of relevant medical information.This is moderately consistent with the reported adverse event/incident.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported as deceased.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.G3: date of received by manufacturer.H6: investigation finding codes - remove code 3233.H6: investigation conclusion codes - remove code 11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15341900
Report Number2031527-2022-00218
Device Sequence Number14596118
Product Code MIH
UDI-Device Identifier00818009015136
UDI-Public(01)00818009015136(17)50125
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/25/2025
Device Model NumberBEA22-90/I16-30
Device Lot Number2543437-012
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 08/16/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date09/01/2022
Supplement Report FDA Received Date11/01/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/22/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX INFRARENAL AORTIC EXTENSION, LOT # 2540228-005.
Outcome Attributed to Adverse Event Death;
Patient Age80 YR
Patient SexMale
-
-