• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MIZUHO OSI TABLE, HANA, 110 V; OPERATING TABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MIZUHO OSI TABLE, HANA, 110 V; OPERATING TABLE Back to Search Results
Model Number 6875
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Codes Damage to Ligament(s) (1952); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/26/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that when the hospital staff were transferring the patient to the hana table for a surgery, the table although locked, moved and they almost dropped the patient.Patient was not injured but one of the employees sustained a wrist injury.
 
Additional Manufacturer Narrative
Preventative maintenance of the table was performed before the incident (b)(6) and it was indicated that the casters had corrosion.Castor replacement was recommended but was not implemented by the hospital at that time.Following the incident (08/26), inspection was done by our field service engineer where the casters of the table were found to be corroded and were replaced as part of the visit.
 
Event or Problem Description
It was reported that when the hospital staff were transferring the patient to the hana table for a surgery, the table although locked, moved and they almost dropped the patient.Patient was not injured but one of the employees sustained a wrist injury.The injured employee was taken to the emergency room for treatment and worker's comp case was opened for this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TABLE, HANA, 110 V
Common Device Name
OPERATING TABLE
Manufacturer (Section D)
MIZUHO OSI
30031 ahern ave
union city CA 94587 1234
Manufacturer (Section G)
MIZUHO OSI
30031 ahern ave
union city CA 94587 1234
Manufacturer Contact
krina shah
30031 ahern avenue
union city, CA 94587-1234
5104291500
MDR Report Key15342068
Report Number2921578-2022-00018
Device Sequence Number14738386
Product Code JEA
UDI-Device Identifier00842430108747
UDI-Public00842430108747
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 10/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6875
Device Catalogue Number6875
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/30/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date09/01/2022
Supplement Report FDA Received Date10/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-