| Catalog Number |
14-521520 |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a distributor received the wrong screw in the package on two screws.The package label is for a fixed screw, but there is a variable screw in the package.There was no patient involvement.This is report one of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report 3012447612-2022-00212.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information in h6: component, investigation type, findings, and conclusions device evaluation: visual inspection revealed the label on the packages state the screw is a fixed screw, part number 14-521520, however the screw in the package is a variable screw, part number 14-521620.Potential cause: root cause is being evaluated per the company's capa process.Dhr review: per dhr review, the dhr does not indicate an issue, however one has been identified per the two complaint files on system.Device use: this device is used for treatment.A follow-up report will be submitted if new information is received that changes the information provided in this report.
|
| |
|
Event or Problem Description
|
|
It was reported that a distributor received the wrong screw in the package on two screws.The package label is for a fixed screw, but there is a variable screw in the package.There was no patient involvement.This is report one of two for this event.
|
| |
|
Search Alerts/Recalls
|