| Catalog Number |
MC1342P |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Without a product return, no product evaluation is able to be conducted.Current information is insufficient to permit a valid conclusion about the cause of this event.If additional information is obtained that adds value to the relevant content of this report and/or a conclusion can be drawn, a follow-up report will be sent.Reference report (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported an roi-c cage and plate were broken during installation within surgery.The removal of these implants caused a delay of 30 minutes but the patient is in good condition after surgery and has been discharged from the hospital.This is report two of two for this event.
|
| |
|
Event or Problem Description
|
|
It was reported an roi-c cage and plate were broken during installation within surgery.The removal of these implants caused a delay of 30 minutes but the patient is in good condition after surgery and has been discharged from the hospital.This is report two of two for this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspection: the products were returned and visual inspection found that the plate has some gouges/deformation damage and the cage is deformed and fractured.Dhr review: the dhrs were reviewed.There are no indications of manufacturing issues which would have contributed to this event and the devices were likely conforming when they left zimvie¿s control.Potential root cause: a definitive root cause cannot be determined with the information provided.This event could possibly be attributed to off-axis forces applied during use.The damage to the returned parts could have also occurred during the removal of the parts.Device usage this device is used for treatment.A follow-up report will be sent if additional information is obtained that adds value to the relevant content of this report.
|
| |
|