| Model Number |
AORTIC CUTTERS, 5-PACK (3.8MM) |
| Medical Device Problem Code |
Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/18/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Trackwise id (b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.Device discarded.
|
| |
|
Event or Problem Description
|
|
The hospital reported that during a coronary artery bypass procedure using aortic cutter (3.8mm), after releasing the safety lock, it was not possible to use it because it was activated just by touching the activation button lightly.There was no effect on the patient because he was wearing a cap.Surgery successfully completed using new ac-3038 cutter.
|
| |
|
Additional Manufacturer Narrative
|
|
Trackwise#: (b)(4).The lot # 25162078 history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|