PHILIPS VOLCANO REFINITY CATHETER, SHORT TIP, 89900; CATHETER, ULTRASOUND, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
89900 |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Codes |
Chest Pain (1776); Restenosis (4576)
|
| Date of Event |
08/13/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This case was reviewed and investigated according to the manufacturer¿s policy.Foreign source: country is denmark.The refinity catheter was not returned for evaluation, thus no returned product investigation was performed.Do not apply to this submission.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|
Event or Problem Description
|
|
It was reported that during a coronary procedure to treat an elevated st segment in the patient''s left anterior descending (lad), a non-manufacturer''s balloon was used for pre-dilatation to establish blood flow from timi 0 to timi 3.Then, a refinity ivus catheter was being prepared for use to measure the diameter of the vessel for stent placement, but displayed no image.The refinity was disconnected and reconnected, but still no image.A second refinity catheter was opened, but also displayed no image, and re-attempted connection without success.Therefore, a third refinity catheter was opened and experienced the same failure.After approximately 15 mins, and while attempting to use the third refinity catheter, the patient complained of chest pain and the ecg showed re-elevation of the st segment.Angiography showed re-occlusion of the vessel.The vessel was pre-dilated and timi 3 flow was re-established.The procedure was completed with a non-manufacturer¿s ivus guided stent.This adverse event and product problem is being submitted to capture the third malfunctioning refinity catheter (not used in patient), which delayed the diagnosis and resulted in re-occlusion of the vessel, requiring intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h3: the refinity catheter was returned for evaluation.Visual inspection found a portion of the drive cable exposed and a portion of the shaft was stretched and broken.No missing material was detected.Image verification test could not be performed due to the damage observed on the catheter.Block h6: the probable cause of the reported failure "no image" could not be conclusively determined since the catheter was received damaged.It is unknown when the damaged occurred; therefore, the cause could not be established.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|
| |
|