| Model Number |
ZXR00 |
| Medical Device Problem Code |
Material Opacification (1426)
|
| Health Effect - Clinical Code |
Discomfort (2330)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: exact date is unknown/not provided.Best estimate is prior to on (b)(6) 2022.Implant date: if implanted, give date: unknown, information was requested but not provided.Explant date- explant date: lens remains implanted, therefore not explanted.The lens was not returned for evaluation as it remains implanted.There was no model or serial number reported for this device; therefore, no further investigation can be performed.If there is any further relevant information received, a supplemental medwatch will be filed.An attempt has been made to obtain additional information regarding the event; however, to date, the information has not been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
A doctor reported that a patient who was implanted with a symfony intraocular lens (iol) last month on (b)(6) 2022 ended up a -1.50.This was the patient's second eye and the calculatios were very similar to when the patient's first eye was done a few years ago by another doctor.The doctor also indicated that the lens has some unusual central haze that is very circular and confined to the first ring, that it is not capsule.The doctor is planning on exchanging the lens being concerned that it could cause future visual issues.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information/corrected data: additional information received from the doctor indicated that patient¿s issue and the central haze with their right eye first noticed three (3) week post operation.It was indicated that the patient does not have debilitating condition.The patient is scheduled for an explant on (b)(6) 2022.The doctor stated that he is exchanging the lens since the haze is central and the patient had a myopic surprise.The dcotor is hoping this myopia is not due to the haze.No further information was provided.The following fields were updated accordingly: section a2: age / date of birth: 81 years / (b)(6) 1941.Section a3: gender: male.Section a5: ethnicity / race: not hispanic/latino / white.Section b3: date of event: 3 week post op visit.Section d1: brand name: tecnis iol.Section d3: manufacturer name and address: amo puerto rico mfg.Inc., road 402 north, anasco ind.Pk, anasco, puerto rico, 00610.Section d3: manufacturer email address: (b)(6).Section d4: model number: zxr00.Section d4: catalog#: zxr00u0220.Section d4: serial number: (b)(6).Section d4: expiration date: sep 11, 2026 section d4: udi number: (b)(4).Section d6a: if implanted, give date: (b)(6) 2022.Section h4: device manufacture date: sep 11, 2021.Section h3 - other (81): the lens was not returned for evaluation as it remains implanted.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: the source photo from account was received and evaluated by the johnson & johnson subject matter expert (sme).A pseudophakic eye was displayed claiming to be implanted with a tecnis symphony erv, intraocular lens, model zxr00.The anterior segment is displayed in the slit lamp picture and a lens with a optical diffractive pattern can be observed.The actual location of the haziness/cloudiness cannot be determined.The root cause and potential clinical impact of the observed phenomena cannot be determined from a picture assessment.The complaint issue was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Therefore, there is no indication of a product deficiency or product malfunction.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|